Measurement and Testing

Dynamic generation of trace gas standard mixtures of both SVOCs and oxygen-sensitive SVOCs at ppm(v/v) to sub-ppb(v/v) concentrations using Tracer Cert® diffusion tubes

Apr 12 2016

Author: John M. Thompson on behalf of Tracer Measurement Systems Ltd

Free to read

This article has been unlocked and is ready to read.

Download

ISO 6145-8-2005 "Gas Analysis - Preparation of calibration gas mixtures using dynamic volumetric methods - Part 8: Diffusion method" shows designs of diffusion tubes/cells that are clearly very difficult to fill. Tracer Measurement Systems has developed a new design of uniquely identifiable, laser-engraved, refillable and re-useable diffusion tubes (see Journal of Environmental Monitoring, 2009, 11, 1543-4) which is much easier to fill and use than the design shown in ISO 6145-8-2005. This presentation shown at the 2015 PEFTEC conference is now available to download.

A novel method of use of these Tracer Cert® diffusion tubes at -21°C in a new Tracer Cert® Sub-ambient Thermostat™ for the dynamic generation of oxygen-sensitive VVOCs, isoprene and acetaldehyde, in an oxygen-free nitrogen carrier gas flow at ppb(v/v) concentrations is described and the results of the calibrations of the diffusion tubes are shown. This method can be widely applied to the dynamic generation of trace standards of oxygen-sensitive VVOCs in the ppm(v/v) to sub-ppb(v/v) concentration range.


Digital Edition

PIN Buyers Guide 24.6

January 2024

In This Edition Product Listings & Exhibition Diary - A-Z of Products and Services - A-Z of Company Contact Details Safety Articles - The technology behind the ION Science Tiger XT -...

View all digital editions

Events

Nepal Lab

Feb 29 2024 Kathmandu, Nepal

IFPAC 2024

Mar 03 2024 Bethesda, MA, USA

ChemTech World Expo 2024

Mar 04 2024 Mumbai, India

SIAF Guangzhou

Mar 04 2024 Guanghzou, China

View all events