Determination of Flame Retardants by Gas Chromatography – Mass Spectrometry

Mass spectrometry & spectroscopy

Determination of Flame Retardants by Gas Chromatography – Mass Spectrometry

03 Aug, 2020

Published over 5 years ago. See the latest and most current information on Mass spectrometry & spectroscopy.

Ashleigh Mellor
5 min read
Download

Polybrominated biphenyls (PBBs) and Polybrominated diphenyl ethers (PBDEs) are routinely used as flame retardants in electronic equipment, textiles and plastics. These flame retardants are deemed a hazardous substance due to the health implications which they cause; they are carcinogenic compounds. PBBs and PBDEs are regularly released into the environment during production and disposal of products which contain these flame retardants. As carcinogenic chemicals, the release of PBBs and PBDEs must be controlled and monitored to minimise the amount released into the environment and limit human exposure. The Restriction of Hazardous Substances Directive (RoHS) regulates the levels permitted for use in electronic equipment, typically 0.1% for both PBBs and PBDEs. SCION Instruments developed a method for the analysis of PBBs and PBDEs by gas chromatography with mass spectrometry.

Experimental

A SCION 456 GC was coupled to the SCION Single Quad Mass Spectrometer. The analytical conditions for this analysis can be found in Table 1 with the instrumentation used being found in Figure 1.

Table 1. Analytical conditions of the GC-MS.

Conditions    

S/SL: 280°C

Column: Scion-5HT 15m x 0.25mm x 0.10μm

Oven: 110°C (2 mins), 40°C/min to 200°C, 10°C/min to 260°C, 20°C/min to 340°C (2 min)

Carrier Gas: Helium 1mL/min constant

Transfer Line: 300°C

Source: 230°C

MS: Full Scan, 100-1000Da

1mL of each PPB and PBDE standard mixtures were prepared in a 5mL volumetric flasks. Toluene was used to adjust the standard stock solutions to 20µg/mL. Working standards were prepared in concentrations of 0.05, 0.15, 0.25, 0.35 and 0.45µg/mL for each of the target compounds. Preparation of samples varies depending on the sample type but can include pyrolysis and solvent extraction.

Results

The five working standards were analysed in both full scan and SIM mode, with quantification using SIM mode. The total ion chromatogram (TIC) for a 20µg/mL stock standard was used for peak identification with peak retention times being compared with the certificate of analysis to confirm identification. Figure 2 shows the two total ion chromatograms (TIC) from the 20µg/mL stock standards whilst Table 2 details peak identification, linear coefficient values of all calibration curves and repeatability values, in full scan mode. Repeatability of the system was determined through ten replicates of each component at 0.05µg/mL.

Table 2. Peak identifiers, retention time, RSD% and linear coefficient values.

Peak    Peak ID    RT (min)    RSD % (RT)    RSD% (Area)    Linear Coefficient(R2)

-    PBBs    -    -    -    -

1    2-bromobiphenyl    3.36    0.02    2.4    0.99992

2    2,5-dobromobiphenyl    4.23    0.01    2.9    0.99992

3    2,4,6-tribromobiphenyl    4.84    0.01    2.3    0.99963

4    2,2’,5,5’-tetrabromobiphenyl    6.01    0.01    2.6    0.99943

5    2,2’,4,5’,6-pentabromobiphenyl    7.01    0.006    1.9    0.99823

6    2,2’,4,4’,6,6’-hexabromobiphenyl    8.10    0.01    2.4    0.99782

7    2,2’,3,4,4’,5,5’-heptabromobiphenyl    11.42    0.007    2.7    0.99714

8    Octabromobiphenyl    12.80    0.006    3.9    0.99525

9    2,2’,3,3’,4,4’,5,5’,6-nonabiphenyl    13.49    0.005    2.1    0.99829

10    Decabromobiphenyl    14.06    0.005    2.1    0.99510

-    PBDE        -        -

11    4-monobromobiphenyl ether    3.77    0.006    2.5    0.99958

12    4,4’-dibromodiphenyl ether    4.71    0.03    2.7    0.99930

13    3,3’,4-tribromodiphenyl ether    5.79    0.007    1.4    0.99855

14    3,3’,4,4’-tetrabromodiphenyl ether    7.54    0.02    4.0    0.99724

15    2,2’,4,4’,6-pentabromodiphenyl ether    8.16    0.01    1.4    0.99588

16    2,2’,4,4’,5,6’-hexabromodiphenyl ether    9.53    0.07    1.7    0.99619

17    2,2’,3,4,4’,5,6;-heptaBDE    11.91    0.01    2.9    0.99957

18    2,2’,3,4,4’,5,5’,6’-octaBDE    12.67    0.07    4.0    0.99889

19    2,2’,3,3’,4,4’,5,5’,6-nonbrominated diphenyl ether    13.84    0.009    2.9    0.99963

20    Decabromodiphenyl ether    14.73    0.0082    3.4    0.99903

It is critical that the instrumentation used can detect both low levels and higher concentrations to determine how much of a risk the flame retardant items are to the environment and human health. Excellent linearity was observed for all target compounds, with a calibration range from 0.05µg/mL to 0.45µg/mL. All coefficients were equal to or greater than 0.995. Figure 3 shows the calibration curve of Decabromobiphenyl, which is representative of all target compounds analysed. Linearity was calculated using SIM data.

In addition, excellent repeatability was also observed for all compounds for both retention time and peak area. These low RSD% values are important when monitoring low-level contaminants, such as PPB and PBDE, as there must be confidence in the levels reported especially when there are health implications associated with the target compounds. As the RoHS limits the exposure of PPBs and PBDEs to no more than 0.1% (by mass), the repeatability of the 0.05µg/mL calibration standard was key to this analysis.

Conclusion

Optimisation of the SCION GC-MS allowed excellent separation and quantification of polybrominated biphenyls and polybrominated diphenyl ethers, common flame retardants. Using a quantitative SIM method it was possible to identify and quantify twenty components in fifteen minutes, over a concentration range of 0.05µg/mL to 0.45µg/mL.

Monitoring these carcinogenic compounds is vital for industries which manufacture electrical and plastic products, in order to reduce the amount released into the environmental during disposal.

PIN 27.2 Apr/May 2026

Explore our Digital Edition

Discover the latest news and research

Digital edition

Explore Our Other Sites

Labmate Online
Pan-European One Health network targets infectious disease threats
Explore more Arrow
Envirotech Online
Rack-mountable FTIR gas analyser for integrated multi-gas analysis in fixed measurement systems
Explore more Arrow
Pollution Solutions Online
Queen’s University Belfast leads £2.9 million international project to combat Malaysia’s growing e-waste crisis
Explore more Arrow
Chromatography Today
Chromatography and XFEL imaging reveal critical point behind water’s behaviour
Explore more Arrow