Analytical Instrumentation

  • Analysis of S, Fe, Ni and V in Residual Oil According to IP610/13
  • Figure 1
  • Table 1

Analysis of S, Fe, Ni and V in Residual Oil According to IP610/13

Nov 17 2020 Read 391 Times

Introduction

Residual oil, which contains sulphur (S), iron (Fe), nickel (Ni), vanadium (V) and other elements, is obtained after distilling crude oil. It is used for marine fuel or electric power plant fuel. When residual oil is burned, some elements can cause damage to marine engines and the environment. For example, sulphur is considered the most harmful element causing air pollution, and metals like Ni, V and Fe are regarded as catalyst poisons, blocking a catalyst’s performance.

X-ray fluorescence (XRF) is a powerful tool for analysing these elements in residual oil. Sample preparation is very simple and analysis results are obtained quickly. This application report describes the analysis of S, Fe, Ni and V by IP Test Method 610/13 - Determination of sulphur, vanadium, nickel, iron, calcium, zinc and phosphorous in residual fuel oil by wavelength dispersive X-ray fluorescence spectrometry – and demonstrates the performance of the Rigaku ZSX Primus IVi sequential wavelength dispersive X-ray fluorescence (WDXRF) spectrometer.

Instrument

The ZSX Primus IVi tube-below WDXRF spectrometer is optimised for the routine analyses that today’s petroleum laboratories need to perform. The optical chamber can be maintained under vacuum while the sample chamber is under helium atmosphere by using a vacuum seal between the optical chamber and the sample chamber. The vacuum seal minimises helium gas consumption and reduces the time required for helium purge. The spectrometer is equipped with a 4 kW X-ray tube with an ultra-thin beryllium window or a 3 kW X-ray tube.

The operation software is designed for easy routine analysis. The quantitative analysis flow-bar guides users step by step to generate applications. The sample ID table and program operation help operators carry out daily analysis.

Sample and sample preparation

Residual oil standard samples supplied by Analytical Services, Inc. were used for calibration.

Eight grams of each residual oil sample was poured into a liquid cell (Chemplex® 1095). 3.6 μm thick Mylar® (Chemplex® 150) was used as the sample film.

Measurement and calibration

Measurements were performed on the ZSX Primus IVi spectrometer with a high-performance 4 kW Rh-target X-ray tube for S, Fe, Ni and V. Primary beam filters were used to protect the X-ray tube window in case of sample leakage. In addition, the filter reduces the background to improve the peak-to-background ratio.

The calibration curves were generated with a matrix correction applied to all the analytes. The correction coefficients used were calculated theoretically by the fundamental parameter (FP) method. The calibration curves are shown in Figure 1.

Analysis results

Typical lower limits of detection (LLD) for Fe, Ni and V calculated using the standard sample with the lowest content.

Repeatability tests for two samples (sample A and B) were carried out for 21 test portions according to “method A” of IP610-13. The test results are tabulated in Table 1, in which the differences between two successive aliquots are shown and “r” represents “repeatability” as defined in IP 610/13, which states that the difference between two successive test results obtained by the same operator with the same apparatus under constant operation conditions on identical test material would, in the normal and correct operation of the test method, exceed the value for repeatability defined in IP 610/13 METHOD A.

The test results shown in Table 1, where the difference between two successive aliquots does not exceed the repeatability (r) for each analyte, prove that the performance of the ZSX Primus IVi spectrometer meets the requirement of ISO610-13 for S, Fe, Ni and V analysis.

Reader comments

Do you like or dislike what you have read? Why not post a comment to tell others / the manufacturer and our Editor what you think. To leave comments please complete the form below. Providing the content is approved, your comment will be on screen in less than 24 hours. Leaving comments on product information and articles can assist with future editorial and article content. Post questions, thoughts or simply whether you like the content.

Post a Comment





Digital Edition

Petro Industry News Buyers Guide 2021

January 2021

In This Issue Fuel For Thought - Signal Group announces new senior appointment - Network of organisations committed to workplace launched - Ayalytical Instruments acquires Lawler Manufactur...

View all digital editions

Events

SLAS 2021

Jan 23 2021 Digital Event

European Gas Conference

Jan 26 2021 Virtual event

EUEC 2021

Feb 08 2021 Virtual & live stream

VIRTUAL METECH 2021

Feb 15 2021 Online

View all events