Analytical Instrumentation

Get More Precision at Trace Level: Sulphur-Analysis in Automotive Fuels According to Iso 20884 and Astm D 2622 By WDXRF

Jan 08 2013

Author: Dr. Kai Behrens, Dr. Arnd Bühler on behalf of Bruker AXS

Free to read

This article has been unlocked and is ready to read.

Download

In the last years fuel quality has been increasingly regulated by legislation to enforce more stringent automotive emission levels. The most important characteristic is the sulphur concentration in different fuel types. The allowable limit of sulphur in automotive fuels went down to the lowest mg/kg range. A limit of 50 mg/kg sulphur was announced in Europe for 2005, but tax incentives for fuels established even lower levels. Since 2009 Euro V Diesel is enforced with a maximum concentration of 10 ppm sulphur. In the US EPA regulations are enforcing a level of less than 15 ppm for Highway Diesel fuel. The preferred technology to analyse low sulphur at this low level is wavelength dispersive X-ray fluorescence spectrometry (WDXRF). But these low levels are no longer covered by ISO 14596 or ASTM D 2622 for high ranges.

The norm which describes the low sulphur fuels analysis by WDXRF is now ISO 20884 and ASTM D 2622 for the very low range down to 5 ppm. As matrix effects hardly vary when analysing fuels, only ISO 20884 and ASTM D 2622 could be established without an external standardisation. The concentration range is subdivided in a low range (5 - 60 mg/kg) and a high range (> 60 - 500 mg/kg). This report describes how the modern WDXRF spectrometer S8 TIGER analyses effectively the low sulphur type fuels on a daily basis. The high spectral resolution and the enhanced light element determination the wavelength dispersive X-ray fluorescence (WDXRF) spectrometer S8 TIGER easily achieves detection limits down to 0.2 ppm.  

Instrument

The S8 TIGER is perfectly suited for the low sulphur analysis in a refinery or a commercial testing lab. Even lowest traces are analysed efficiently and reliably based on the optimum instrument setup. With excitation power of up to 4 kW, a high resolution, high intensity optimised analyser crystal XS-GE-C the S8 TIGER provides optimal analytical performance. The system is optimised for the analysis of liquid samples to make the operation as simple and failsafe as possible: 

Liquid samples are automatically detected during the loading and the helium mode is enforced to prevent spillage of the sample and damages on system components. The low temperature X-ray tube head and the unique atmospheric helium mode prevent volatile samples from boiling which finally protects the instrument and ensures the best analytical stability. Finally the SampleCare system ensures with the unique vacuum seal the protection of spectrometer components separating the sample and spectrometer chamber. Fumes and droplets will not enter the spectrometer chamber. This completely protects crystals and detectors versus against damage and avoids frequent system failures like with conventional spectrometers on the market. 

The S8 TIGER comes for ISO 20884 and ASTM D 2622 with PETRO-QUANT and the ready-to-analyse solution containing the calibration inclusively the specific set of calibration standards, the optimised measurement method and drift correction samples. For instrument verification and the performance test in addition a quality check sample is supplied with the package. This helps to establish easily the analytical quality routine for audit conformity.

Sample preparation and measurement parameters

7 g of the gasoline sample are filled in a liquid cup with a 3.6 µm Mylar foil. This foil is transparent for the light element radiation, but provides chemical resistance against gasoline.

All data obtained using the following measurement parameters are listed in table 1. The helium mode with atmospheric pressure is applied, because of the high volatility of fuel samples.

Calibration of the Low Concentrations Range

The norm compliant calibration is based 5 standards ranging from 5 to 50 ppm plus blank sample, and is based on a linear model. The following graph and table describes the calibration (concentrations in mg/kg):

The mean regression deviation of the calibration is less than 0.1 ppm, the squared correlation coefficient is with a value of 0.999957 very close to 1. The detection limit LOD (3s, 24 seconds) is 0.4 ppm.

Repeatability Low Concentration Range

The repeatability of the ISO 20884 method was checked with a 10.3 ppm QC-sample by measuring the sample 22 times. According to ISO 20884, the difference between two consecutive results in the range of 10.3 mg/kg must not exceed 1.9 mg/kg in more than one out of 20 cases.  The maximum deviation of two consecutive measurements was 0.9 ppm, finally much less, the relative standard deviation was 0.3 ppm. These results are fully compliant with ISO 20884.

Calibration of the High Concentrations Range

The norm compliant calibration is based 6 standards ranging from 50 to 500 ppm plus blank sample and is based on a linear model. The following graph and table describes the calibration (concentrations in mg/kg):

The mean regression deviation of the calibration is less than 0.8 ppm, the squared correlation coefficient is with a value of 0.999963 very close to 1. The detection limit LOD (3s, 30 seconds) is 0.4 ppm.

Repeatability Low Concentration Range

The repeatability of the ISO 20884 method was checked with a 150 ppm QC-sample by measuring the sample 22 times. According to ISO 20884, the difference between two consecutive results in the range of 10.3 mg/kg must not exceed 4 mg/kg in more than one out of 20 cases.  The maximum deviation of two consecutive measurements was 1.4 ppm, finally much less than described in the standards as maximum difference between two consecutive measurements. The relative standard deviation was 0.5 ppm. These results are fully compliant with ISO 20884 and ASTM D 2622.

Results and Conclusion

The S8 TIGER with 4 kW excitation power and the curved germanium analyser crystal XS-GE-C easily achieves the analytical performance required for the determination of low sulphur in automotive fuels according to ISO 20884 and ASTM D 2622. It delivers impressive analytical precision allowing a close monitoring of the product quality. The requirements of ISO and ASTM are easily achieved. The results are shown in Table 6.

Therefore the S8 TIGER is perfectly suited for the sulphur test of automotive fuels in refineries, commercial testing labs and for governmental service labs in daily routine. Providing best analytical performance in combination with simple intuitive operation and failsafe handling of liquid sample the S8 TIGER provides peace of mind. The PETRO-QUANT solution package for ISO 20884 and ASTM D 2622 even provides the audit conform long term performance test, so even analytical quality audits are easy to pass.

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

PIN 25.2 Apr/May

April 2024

In this Edition Safety - Carbon monoxide toxic and flammable gas detection Analytical Instrumentation - Density: A fundamental parameter at critical stages within the petroleum sector...

View all digital editions

Events

Saudi Plastics and Petrochem

May 06 2024 Riyadh, Saudi Arabia

OTC 2024

May 06 2024 Houston, Tx, USA

ILTA 2024

May 06 2024 Houston, Tx, USA

Canada Gas & LNG Exhibition & Conference

May 07 2024 Vancouver, BC, Canada

POGEE

May 08 2024 Lahore, Pakistan

View all events