Ultimate Sealing Coulometric Titration Vessel

Apr 07 2015

There are various designs of titration vessel used on coulometric Karl Fischer titrators and they all have one thing in common, the interior of the titration vessel should be sealed from ingress of atmospheric moisture.

To achieve this, the most frequently used design is to have ground glass joints on the generator and detector electrodes which are then fitted into ground glass joints on the lid of the titration vessel. However it is well known that dry ground glass joints will have a tendency to stick to each other and then the problems begin.

The most expensive part of any coulometric Karl Fischer instrument is generally the glassware, especially the titration vessel and electrodes. If these become stuck then it can be a very costly exercise trying to separate them, often resulting in breakages.

To overcome this problem many manufacturers recommend smearing grease on the electrode joints or, alternatively, fitting PTFE sleeves over the joints. Each of these methods are equally effective but they are also equally problematic. If grease is being used to seal the electrodes into place then it is necessary to turn the joints at least once per week to ensure that the grease has not dried out and the electrodes have become stuck. This precaution is usually stated clearly in the user manual. Similarly, when the glassware is periodically cleaned and charged with fresh reagents, the PTFE sleeves should also be cleaned, dried and refitted to the electrode joints. These sleeves are often lost or damaged which, once again, leads to electrodes being stuck in the titration vessel.

Originally designed for coulometer users on North Sea oil and gas platforms, the GR Scientific Low Drift Cell (LDC) uses a special design of glass joint which will not stick or jam and is totally grease and sleeve free. The connecting cap of the electrode is screwed down to compress the O-ring on to the top of the female joint which is part of the titration vessel, ensuring a perfect seal. The electrode is released with ease by simply unscrewing the connecting cap which engages the loosening ring to lift the electrode free from the joint.


Digital Edition

Lab Asia 31.2 April 2024

April 2024

In This Edition Chromatography Articles - Approaches to troubleshooting an SPE method for the analysis of oligonucleotides (pt i) - High-precision liquid flow processes demand full fluidic c...

View all digital editions

Events

SETAC Europe

May 05 2024 Seville, Spain

InformEx Zone at CPhl North America

May 07 2024 Pennsylvania, PA, USA

ISHM 2024

May 14 2024 Oklahoma City, OK, USA

ChemUK 2024

May 15 2024 Birmingham, UK

Water Expo Nigeria 2024

May 21 2024 Lagos, Nigeria

View all events