Analytical Instrumentation

Fast and Reliable Chlorine Determination by Self-Adopting Systems

Dec 10 2015

Author: Dr. Stefan Jezierski and Dr. Angela Gröbel on behalf of Analytik Jena GmbH

Free to read

This article has been unlocked and is ready to read.

Download

During storage, transportation and further processing, chlorine containing compounds can form harmful substances (HCl, metal halides …), which consume costly parts of the facilities (e.g. tanks, pipelines, catalysts). To avoid this, and to ensure a constant high quality and yield of the final products, it is crucial to control the chlorine content of the feeds and different process streams permanently. Therefore reliable results must be available in next to no time. This task is a real challenge for common techniques, since they are either limited to special matrix types or narrow concentration ranges. That’s why numerous diverse techniques must be applied to cover the full range. This occupies not only manpower but also valuable time. A new generation of intelligent self-adapting analysers can produce the relief, which is urgently needed.

Chlorine is Everywhere

Chlorine is present everywhere in the world, whether organic or inorganic bound. Especially in refineries and petrochemical industry such materials can cause serious problems in numerous processes. This is the reason for monitoring chlorine constantly in feed materials, intermediates, final products, and waste and by-products. If the specified maximum permissible values are exceeded, not only the efficiency of plant will decrease but also the maintenance effort and costs (e.g. catalyst poisoning) will rise and the entire facility may suffer from considerable damages by corrosion and wearing.
In order to prevent such effects, a precise and quick determination is important. Depending on the matrices which have to be determined resp. the expected concentration range, many different, more or less automatable techniques were applied. Besides whet chemical methods coupled with ISE, photometry etc., and XRF, mainly combustion techniques (e.g. bomb, Schöninger) are in use. Among all these techniques, the high temperature combustion coupled with micro coulometric titration has proved to be most suitable. It delivers not only fast analysis and high sensitivity, but also a maximum degree of automation.

Optimum Combustion is the Base

As much as the concentrations differ (ppb – wt-%), as much the combustion behaviour of the varied matrices differs. Their volatility, boiling intervals and pyrolysis tendency are variable in a wide range. When analysing just one matrix type, this might not be a big issue, but for analysis of varied matrices manifold problems do appear.
The biggest challenge for sure is to create optimum conditions for sample digestion, and thereby, complete conversion of all sample components. Only this way it is possible to generate trustable results. Even the smallest amount of soot or cracking products will affect final results considerably. To avoid this, either plenty of time and experienced operators to create suitable combustion methods for each sample matrix are necessary, or just one auto-adaptive combustion system, which can be operated by everyone. Such optimisation can be realised by the flame sensor technology. It automatically adapts the combustion process to the special needs of each sample component and matrix type. Thus anybody can realise a quantitative combustion without any effort, independently from used sample quantities.

Table 1. Effects of different digestion modes on recovery of 100 µl 500 µg/l Cl reference matrix (Download the article to view this table)

The manual or automatic fast introduction at constant rate (CR boat) cause incomplete combustion, low recovery and thus bad repeatability of the results. On contrary, low RSD and better recovery can be achieved using an adopted boat program (including waiting points and different rates), however very long analysis times are crucial to realise this. The flame sensor technology optimises introduction time and helps to achieve satisfying recovery and RSD at once.

Experimental

The samples were analysed without pretreatment using the multi EA® 5000 Cl in horizontal mode. The liquid and solid samples were dosed by means of a multi matrix sampler. Afterwards the transfer into the furnace was carried out fully automatically by means of a flame sensor controlled automatic boat drive. The combustion takes place in a biphasic process. During the first phase volatile components were evaporated in an inert gas atmosphere and purged into the combustion zone, to be oxidised completely at 1050°C. In the second step remaining heavier sample components and possibly formed pyrolysis products, were combusted in pure oxygen. Before they were guided to the coulometer cell, the reaction gases, including HCl, are cleaned and dried by the auto-protection system and sulfuric acid. Meanwhile a powerful heating avoids condensation, which would affect Cl recovery remarkably. When HCl has arrived, it is absorbed in the cell electrolyte and titrated subsequently. An integrated cell cooling, drift control and usage of lightproof material, ensure a stable performance. This enables precise analysis results even for the lowest trace contents.

 

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

PIN 25.1 Feb/March

March 2024

In This Edition Safety - The technology behind the ION Science Tiger XT - Safety with ammonia and LOHCs as hydrogen carriers Analytical Instrumentation - Discussion on new tribology te...

View all digital editions

Events

ENVEX 2024

May 03 2024 Seoul, South Korea

SETAC Europe

May 05 2024 Seville, Spain

Saudi Plastics and Petrochem

May 06 2024 Riyadh, Saudi Arabia

OTC 2024

May 06 2024 Houston, Tx, USA

ILTA 2024

May 06 2024 Houston, Tx, USA

View all events