• PhoCheck+ Triumphs Over Its Competition

Measurement and Testing

PhoCheck+ Triumphs Over Its Competition

Jan 10 2009

The Ion Science PhoCheck+ has triumphed over its closest competitors after a stringent PID testing and selection process by the Maine Department of Environmental Protection (ME DEP).

The evaluation process consisted of both laboratory and field components. The lab tests used an ME DEP developed ‘gas train’ to simulate soil contaminated with fuel oil or gasoline, allowing the ME DEP to set the contamination and vary the humidity. Zero air was then used to adjust the concentrations.

The field test involved using instruments on remediation sites for bag headspace analysis. Several instruments were found to be unsuitable field units for varying reasons including unstable readings, slow response and limited range (less than 2000 ppm).

After initial testing, the PhoCheck+ 3000Ex was further evaluated for ease of use, speed, repeatability and linearity between concentrations. The instrument was proven to give more consistent, clear readings and faster to read results. Maine has now selected PhoCheck+ 3000Ex as the most suitable PID instrument to meet their requirements.

Like all Ion Science PIDs, PhoCheck+ incorporates Ion Science patented Fence Electrode Technology; a unique three electrode format enabling
increased resistance to both humidity and contamination.


Digital Edition

PIN 25.1 Feb/March

March 2024

In This Edition Safety - The technology behind the ION Science Tiger XT - Safety with ammonia and LOHCs as hydrogen carriers Analytical Instrumentation - Discussion on new tribology te...

View all digital editions

Events

AOCS Annual Meeting & Expo

Apr 28 2024 Montreal, Quebec, Canada

The Safety & Health Event

Apr 30 2024 Birmingham, UK

ENVEX 2024

May 03 2024 Seoul, South Korea

SETAC Europe

May 05 2024 Seville, Spain

Saudi Plastics and Petrochem

May 06 2024 Riyadh, Saudi Arabia

View all events