On-Site Measurement of Hydrocarbons in Drill Cuttings

Apr 02 2014 Read 1490 Times

Portable infrared analysers, such as the InfraCal Analyzers from Wilks (USA), have been used worldwide for measuring offshore oil levels in produced water for over 45 years. They have the advantage over other analytical instruments in that the measurement can be performed by non-technical users and it is relatively unaffected by composition changes in the effluent. They also provide offshore users with the necessary ruggedness and reliability that is a key requirement.

In addition to their use for measuring oil content in produced water, portable IR analysers are ideal for both onshore and offshore oil content measurements in drill cuttings. If drill cuttings are to be disposed of via land application or overboard, there are limits on hydrocarbon levels depending on the disposal choice.

While water samples typically have hydrocarbon levels in the part per million (ppm) range, drill cuttings can be in the 1-10% range. Both the InfraCal TOG/TPH Analyzer Model HATR-T2 and the new InfraCal 2 Model ATR-SP are configured with horizontal ATR sample stages. Hexane, pentane or cyclohexane is used as the extraction solvent for oil in water levels from 0.3 to 1000 ppm range. The solvent is evaporated off and the infrared absorbance due to the residual oil film is measured.

 The same InfraCal Analyzers with an ATR sample stage can be used for the higher 1-10% range found in drill cuttings. The sample is diluted by using a solvent that does not have an infrared absorbance at the hydrocarbon wavelength such as tetrachloroethylene (perchloroethylene) or S-316 (dimer/trimer of chlorotrifluorethylene) and, therefore, does not require evaporation. The solvent dilution allows for the high percent range measurements with the same ATR sample stage used to measure the ppm levels.

Reader comments

Do you like or dislike what you have read? Why not post a comment to tell others / the manufacturer and our Editor what you think. To leave comments please complete the form below. Providing the content is approved, your comment will be on screen in less than 24 hours. Leaving comments on product information and articles can assist with future editorial and article content. Post questions, thoughts or simply whether you like the content.

Post a Comment





Digital Edition

Petro Industry News November 2020

November 2020

In This Edition Fuel For Thought - Special offer for computer program for chemical thermodynamic and energy release evaluation - Acquisition to combine French expertise for moisture analysis...

View all digital editions

Events

Interlab Africa

Dec 05 2020 Algiers, Algeria

ASTM - D02 Meeting - CANCELLED

Dec 06 2020 Austin, TX, USA

Biogas Convention 2020

Dec 08 2020 Virtual event

CISILE 2020 - NEW DATES

Dec 08 2020 Beijing, China

View all events