So that you get hot water well used and your money doesn`t vaporize

Jul 03 2014

The use of steam and hot water is essential for the operation of many types of installations and facilities in crude oil refinement. Significant cost savings can be attained through the reuse of the return condensate or steam condensate. However, the purity of this return condensate is of great importance. Contaminated steam or water may lead to corrosion in pipes, boilers and heat exchangers. If a heat exchanger corrodes and is no longer impervious, then mineral oils can leak into and contaminate the return condensate. Therefore, an oil-in-water analyser is required and in this case a TOC analyser is an excellent choice, measuring any oil type in a range from 10 ppb to over 100 ppm, as well as organics that do not belong to oil groups. At refineries the levels that need to be monitored and their alarm thresholds are about forty times cleaner than drinking water (100..1,000 ppb). Therefore, the analytical performance of the TOC analyser must be outstanding meeting tough requirements by the industry regarding accuracy, response times and traceability.

The LAR QuickTOCpurity analyser consists of sample transfer and injection system, reactor, carrier gas circuit, gas detector and calibration system. At the heart of the TOC analyser is the well designed multi-loop injection system as the pre-eminent methodology for sample transference and injection of oily samples.

The second important analyser building block is the ultra high temperature reactor at 1,200°C. The sample, containing hydrocarbon molecules, is injected into the reactor tube. Without exception, all hydrocarbon compounds are completely oxidised, forming carbon dioxide gas. A carrier gas flows through the reactor and transfers this CO2 gas to a non-dispersive infrared detector (NDIR detector). The amount of carbon dioxide that is measured, is linearly proportional to the amount of organic carbon molecules (TOC) in the sample. At 1,200°C it is physically impossible for carbon compounds to stay. Experience demonstrates a full recovery rate even for very hard to oxidise compounds e.g. fluorocarbons, chlorinated hydrocarbons, amines.
The third important analyser building block is the calibration and validation system. The QuickTOCpurity offers a unique patented gas validation feature. Whenever needed, the operator can easily and immediately calibrate/ validate the analyser by using a cylinder filled with a certified gas. This certified gas is very stable and can be stored over a period of several years, making it readily available and always ready for use. The injection loop is filled with the calibration gas, such as a nitrogen mixture, and the content is injected into the reactor. There the gas is oxidised which is then measured by the NDIR detector. Within minutes the gas calibration routine is completed. It is simple, fast and accurate.

Last but not least an important feature of the LAR QuickTOCpurity is the possibility to analyse grab samples from individual heat exchangers. This can save a lot of time in identifying which specific heat exchanger is leaking.
The quick and reliable QuickTOCpurity analyser offers operators the assurance that their analysers are giving accurate results and that they are not pouring expensive condensate down the drain.


Digital Edition

PIN 25.1 Feb/March

March 2024

In This Edition Safety - The technology behind the ION Science Tiger XT - Safety with ammonia and LOHCs as hydrogen carriers Analytical Instrumentation - Discussion on new tribology te...

View all digital editions

Events

AOCS Annual Meeting & Expo

Apr 28 2024 Montreal, Quebec, Canada

The Safety & Health Event

Apr 30 2024 Birmingham, UK

ENVEX 2024

May 03 2024 Seoul, South Korea

SETAC Europe

May 05 2024 Seville, Spain

Saudi Plastics and Petrochem

May 06 2024 Riyadh, Saudi Arabia

View all events