Automatic Determination of the Hydroxyl Number (Hn) According to ASTM E 1899-08 and Din 53240-2

Oct 12 2015

Metrohm (Switzerland) presents a fully automated titrimetric method for the determination of the hydroxyl number according to ASTM E 1899-08 and DIN 53240-2. The hydroxyl number is defined as the mg of KOH equivalent to the hydroxyl content of 1 g of sample. The system used to carry out the test consists of: 814 USB Sample Processor, 905 Titrando, 800 Dosino, Solvotrode and the tiamo software.

The method suggested in ASTM E 1899-08 is based on the reaction of the primary and secondary hydroxyl groups with excess toluene-4-sulfonyl-isocyanate (TSI) to form an acidic carbamate. The latter can then be titrated in a non-aqueous medium with the strong base tetrabutyl-ammonium hydroxide (TBAOH).

The method suggested in DIN 53240-2 is based on the catalysed acetylation of the hydroxyl group. After hydrolysis of the intermediate, the remaining acetic acid is titrated in a non-aqueous medium with alcoholic KOH solution.

The two methods allow, for example, the determination of polyols and oxooils without any tedious and time consuming sample preparation and is therefore a big benefit for laboratories that have to analyse a great number of these samples per day.


Digital Edition

PIN 25.2 Apr/May

May 2024

Safety - Carbon monoxide toxic and flammable gas detection Analytical Instrumentation - Density: A fundamental parameter at critical stages within the petroleum sector - Advancements and...

View all digital editions

Events

AIHA Connect 2024

May 20 2024 Columbus, OH, USA

MPGC

May 20 2024 Dubai, United Arab Emirates

HEFC 2024

May 23 2024 Beijing, China

NGVS 2024

May 23 2024 Beijing, China

Analiza

May 28 2024 Tel Aviv, Israel

View all events